Search Warning Letters

Tuesday, August 31, 2010

iVu Imaging Corporation 8/31/10

  

Department of Health and Human Services logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 Dallas District
4040 North Central Expressway
Dallas, Texas 75204-3128
 

August 31, 2010


Ref: 2010-DAL-WL-14


WARNING LETTER


CERTIFIED MAIL
RETURN RECEIPT REQUESTED


Mark L. Stribling, President and CEO
iVu Imaging Corporation
4121 Grace Lane
Grapevine, Texas 76051


Dear Mr. Stribling:


During an inspection of your firm located at the above-referenced address on February 24 and 26, 2010, March 29 through 31, 2010, and April 1, 2010, investigators from the United States Food and Drug Administration (FDA or Agency) determined that your firm manufactures Sofia™ Automated Tomographic Ultrasound (ATUS) Imaging System. Under Section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321 (h), this product is a medical device because it is intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, or are intended to affect the structure or any function of the body.


This inspection revealed that these devices are adulterated within the meaning of section 501 (h) of the Act, 21 U.S.C. § 351 (h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation found at Title 21, Code of Federal Regulations) (CFR), Part 820.


We received your firm's response from Mr. Mark R. Rogers, Vice President of Operations, dated April 8, 2010, to our investigators' inspectional observations. Mr. Rogers responded that the inspectional observations identified GMP nonconformances in your firm's manufacturing process and that your management intended to complete corrective actions within 90 days. We acknowledge your firm's commitment to meeting all regulatory requirements established by the FDA for medical device manufacturers. However, we consider your firm's response incomplete as your firm's response has neither explained nor provided specific corrective actions to resolve each of the noted inspectional observations. We have not received your further response after 90 days. Please provide us an update of your firm's corrective actions and your accomplishment since the April 8, 2010 response. The Agency may conduct follow-up inspections to assure that your firm's corrections are adequate.


At the conclusion of the Inspection, the investigators issued to you the Form FDA-483 (List of Inspectional Observations) which identifies a number of significant violations including, but not limited to, those described below.


Quality System Violations


1. Failure to establish and maintain procedures for validating the device design in order to ensure that the devices conform to defined user needs and intended uses. The design validation must include testing of production units under actual or simulated use conditions, and risk analysis, where appropriate, and the design validation results must be documented, as required by 21 C.F.R. § 820.30(g). FDA 483 Items 3, 4, and 5. Specifically:


a. Several validation (functional) tests in Design Phase IV were not completed (i.e., (b)(4) Test, (b)(4) Test (b)(4) and (b)(4) Tests, (b)(4) Test, and (b)(4) Test).


b. The (b)(4) Test did not specify specific criteria for accepting or rejecting ((b)(4) phantom breast images acquired. This test documented that an acquisition of ((b)(4)) frames (images) was successful. It did not define what "successful" means or the quality of the images to detect and differentiate hyper and hypo echoic lesions.


c. The (b)(4) Test was conducted on a phantom breast of a nominal size. The test protocol did not explain why other cup sizes were not tested in order to access the ability and limitation of the ultrasound system to detect and differentiate lesions in the smallest and largest cup sizes.


d. The design was not validated using production units or their equivalents. The test report (No. 3135042DAL-001) indicated that electrical testing was conducted on a prototype unit on or before April 30, 2008. Your firm stated to the investigator that the manufacturing process has not significantly changed from the prototype. unit to the current production units. Your firm could not provide supporting documentation to demonstrate that the design and production process and specifications of the prototype unit are equivalent to that of the subsequent production units.


2. Failure to establish and maintain procedures to ensure that the device history records (DHR) for each batch, lot, or units are maintained to demonstrate that the device is manufactured in accordance with the device master record, as required by 21 C.F.R. § 820.184. FDA 483 Item 1(a) and (d). Specifically:


a. Your firm did not verify device serial numbers for accuracy and maintain copies of the product labels in the device history records (DHR) before distributing the devices. For example, the performance check records documented the same serial #(b)(4) for the two separate ultrasound imaging systems installed at two separate user sites on January 29, 2009 and May 4, 2009, respectively.

b. Your firm did not maintain in the performance check records (DHR) all (b)(4) images acquired and tested during the final assembly and testing of the ultrasound imaging systems as required by the Performance Check Procedure, FCD-0158, Revision A.


3. Failure to document the results of installation, inspection, and any required testing in accordance with the manufacturer's instructions and procedures to demonstrate proper installation of the device, as required by 21 C.F.R. § 820.170(b). FDA 483 Item 1(e). Specifically:


Your firm's installation checklist did not document nor attach the results of installation of the ultrasound system units installed at the user sites on August 26, 2008, October 15, 2008, January 29, 2009, and May 4, 2009.


4. Failure to establish and maintain device master records (DMRs) to include or refer to the location of device specifications, production process specifications, quality assurance procedures, and packaging and labeling specifications, and to ensure that each DMR is prepared and approved in accordance with 21 C.F.R. § 820.40, as required by 21 C.F.R. § 820.181. FDA 483 Item 6. Specifically:


Your firm's device master record (DMR) was not complete at the time of the inspection. Your firm confirmed to the investigators that it had a draft DMR and was still working with the contract manufacturer to develop additional  acceptance procedures for the transducer, ultrasound console unit, motor assembly, and software.


Responding to This Warning Letter


You should take prompt action to correct the violations addressed in this letter. Failure to promptly correct these violations may result in enforcement action being initiated by the FDA without further notice. These actions include, but are not limited to, seizure, injunction, and/or civil money penalties.


Federal agencies are advised of the issuance of all Warning Letters about devices so that they may take this information into account when considering the award of contracts. Additionally, pre-market approval applications for Class III devices to which the Quality System regulation (21 CFR Part 820) deviations are reasonably related will not be approved until the violations related to the subject devices have been corrected.


Please notify this office in writing within fifteen (15) working days from the date you receive this letter of the specific steps you have taken to correct the noted violations, including and explanation of how you plan to prevent these violations, or similar violations, from occurring again. You should include documentation of the corrective action you have taken. If your planned corrections will occur over time, please include a timetable for implementation of those corrections. If corrective action cannot be completed within 15 working days, state the reason for the delay and the time within which the correction will be completed.


Your response should be sent to Thao Ta, Compliance Officer, Dallas District Office, Food and Drug Administration, HFR-SW140, 4040 North Central Expressway, Suite 300, Dallas, Texas 75204. If you have any questions about the content of this letter, please contact Mr. Ta at 214-253-5217.


Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your facility. It is your responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the FDA 483 issued at the closeout of the inspection may be symptomatic of serious problems in your firm's manufacturing and quality assurance systems.


You should investigate and determine the causes of the violations, and take prompt actions to correct the violations and to bring your products into compliance.


Sincerely,
/S/

Reynaldo R. Rodriguez, Jr.
Dallas District Director
 

-

Monday, August 30, 2010

Unlimited Nutrition 8/30/10

  

Department of Health and Human Services logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 Atlanta District Office
60 Eighth Street N.E.
Atlanta, GA 30309 

August 30, 2010
 

WARNING LETTER
10-ATL-19


HAND DELIVERED


Mr. Mike McCandless, Owner
Unlimited Nutrition
Scivation, Inc., President
1130 Cherry Lane
Graham, NC 27253


Dear Mr. McCandless:

On February 16 - 17, 2010, the U.S. Food and Drug Administration (FDA) conducted an inspection of your facility located at 1130 Cherry Lane, Graham, NC 27253. We have also reviewed your firm's website, www.smartpowders.com.


This letter concerns your firm's marketing of the following products: "Smart Powders Piracetam," "Primaforce Piracetam," "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore." These products are marketed in violation of the Federal Food, Drug, and Cosmetic Act (the Act) as described below.


Piracetam Containing Products


Your firm markets your piracetam products, "Smart Powders Piracetam" and "Primaforce Piracetam" as dietary supplements; however, both products are excluded from the definition of a "dietary supplement" under section 201(ff)(1) of the Act, 21 U.S.C. § 321(ff)(1). To be a dietary supplement a product must, among other things, "bear[ ] or contain[ ] one or more ... dietary ingredients" as defined in section 201(ff)(1) of the Act. Section 201(ff)(1) of the Act defines "dietary ingredient" as a vitamin, mineral, amino acid, herb or other botanical, or dietary substance for use by man to supplement the diet by increasing the total dietary intake, or a concentrate, metabolite, constituent, extract or combination of any dietary ingredient from the preceding categories. The only substance listed as a dietary ingredient on the labeling for your "Smart Powders Piracetam" and "Primaforce Piracetam" products is piracetam. Piracetam is not a vitamin, mineral, amino acid, herb or other botanical, or dietary substance for use by man to supplement the diet by increasing the total dietary intake. Further, piracetam is not a concentrate, metabolite, constituent, extract or combination of any such dietary ingredient. Thus, because your "Smart Powders Piracetam" and "Primaforce Piracetam" products do not bear or contain any dietary ingredients as defined in section 201 (ff)(1) of the Act, these products do not qualify as dietary supplements under section 201(ff) of the Act.1


Your website and labeling include statements such as the following:


Smart Powders Piracetam


• "Piracetam supports memory and concentration, overall well-being, cardiovascular health, and helps reduce stress and fatigue."


Primaforce Piracetam


• "Piracetam supports memory and concentration, overall well-being, cardiovascular health, and helps reduce stress and fatigue."


The claims listed above make clear that "Smart Powders Piracetam" and "Primaforce Piracetam," are intended to affect the structure or any function of the body of man or other animals. Accordingly, these ·products are drugs, under section 201(g)(1)(C) of the Act, 21 U.S.C. § 321(g)(1)(C), because they are not foods and they are intended to affect the structure or any function of the body. Moreover, these products are new drugs as defined by section 201(p) of the Act, 21 U.S.C. § 321(p), because they are not generally recognized as safe and effective for use under the conditions prescribed, recommended, or suggested in their labeling.


Under sections 301(d) and 505(a) of the Act, 21 U.S.C. § 331(d) and 355(a), a new drug may not be introduced or delivered for introduction into interstate commerce unless an FDA approved application is in effect for it. There are no approved applications for "Smart Powders Piracetam"
and "Primaforce Piracetam." Your sale of these products without approved applications violates these provisions of the Act.


Body Building and Sport Performance Products


In addition, your firm's body building and sport performance products "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore" are unapproved and misbranded drugs under the Act.


Your website, www.smartpowders.com. states that your products contain the following ingredients:


• Advanced Muscle Science Arom-X: 1,4,6-etioallocholan-dione
• Advanced Muscle Science 4-AD UTT: 3,17-keto-etiochol-triene
• G.E.T ArimaDex: 3,17-keto-etiochol-triene
• iForce Nutrition Reversitol: 6-Etioallochol-1 ,4-Diene-3,17-Dione
• Fizogen Off Cycle II Hardcore: 3 17 keto etiochol triene


The above-listed ingredient names are all synonyms for the same ingredient, commonly known as "ATD." "ATD" is an aromatase inhibitor. FDA is unaware of evidence that ATD occurs in vivo in humans or animals.


Your firm markets "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore" as dietary supplements. To be a dietary supplement a product must, among other things, "bear[ ] or contain[ ] one or more ... dietary ingredients" as defined in section 201(ff)(1) of the Act. Section 201(ff)(1) of the Act defines "dietary ingredient" as a vitamin, mineral, amino acid, herb or other botanical, or dietary substance for use by man to supplement the diet by increasing the total dietary intake, or a concentrate, metabolite, constituent, extract or combination of any dietary ingredient from the preceding categories. ATD is not a vitamin, mineral, amino acid, herb or other botanical, or dietary substance for use by man to supplement the diet by increasing the total dietary intake. Further, ATD is not a concentrate, metabolite, constituent, extract or combination of any such dietary ingredient. Therefore, ATD is not a dietary ingredient as defined in section 201 (ff)(1) of the Act, 21 U.S.C. § 321(ff)(1).


Your website includes statements such as the following:


Advanced Muscle Science Arom-X


• "Anti-estrogen complex."
• ''Natural testosterone booster."
• "Libido enhancer."
• "Arom-X is scientifically formulated to suppress estrogen production and restore natural testosterone output while acting as a strong libido enhancer."


Advanced Muscle Science 4-AD UTT


• "4-AD UTT is a unique anabolic solution pro-hormone that converts at a high rate of testosterone (sic)."
• "[W]ill give you better strength and size increases than the old testosterone precursors."


G.E.T ArimaDex


• "A proprietary blend of 7 ingredients including an estrogen blocker ... that have all been shown to increase or maintain testosterone levels."


iForce Nutrition Reversitol


• "By regulating the production and levels of Testosterone, Estrogen, LH, and Cortisol an athlete will ensure PEAK. Performance Reversitol promotes the optimal Hormone Regulation for applying Maximum FORCE!"
• "Usage for promoting hormonal regulation ...."


Fizogen Off Cycle II Hardcore


• "Off Cycle" and "If you are subject to performance enhancing drug testing, do not use this product unless cleared by your sanctioning body"


The statements above make clear that "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore" are intended to affect the structure or function of the body. Accordingly, these products are drugs under section 201(g)(1)(C) of the Act, 21 U.S.C. § 321(g)(1)(C).


As mentioned above, your firm markets "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore" as dietary supplements. Under section 201(g)(1) (last sentence), the structure/function claims made for a dietary supplement must be made in accordance with section 403(r)(6) of the Act, 21 U.S.C. § 343(r)(6), or the product is subject to regulation as a drug. Section 403(r)(6) of the Act, 21 U.S.C. § 343(r)(6), authorizes claims that describe the role of a nutrient or dietary ingredient intended to affect the structure or function of the body, or that characterize the way in which a nutrient or dietary ingredient maintains the structure or function of the body.


However, the claims quoted above for your body building and sport performance products, "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore," do not describe the effects of nutrients or dietary ingredients in the products. Rather, the claims made for each product are made for the product as a whole and relate to its "ATD" content. Since "ATD" is not a nutrient or dietary ingredient, claims about improvement of the structure or function of the body do not conform to section 403(r)(6) of the Act, 21 U.S.C. § 343(r)(6). Accordingly, the claims for "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore" render them drugs within the meaning of section 201(g)(1)(C) of the Act, 21 U.S.C. § 321(g)(1)(C).


Moreover, the above-listed body building and sport performance products are "new drugs," as defined by 201(p) of the Act, 21 U.S.C. § 321(p), because they are not generally recognized as safe and effective for use under the conditions prescribed, recommended, or suggested in their labeling. Under sections 301(d) and 505(a) of the Act, 21 U.S.C. §§ 331(d) and 355(a), a new drug may not be introduced or delivered for introduction into interstate commerce unless an FDA-approved application is in effect for it. Your sale of "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore" without approved applications violates these provisions of the Act.


Furthermore, "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore" are prescription drugs as defined at section 503(b)(1)(A) of the Act, 21 U.S.C. § 353(b)(1)(A), because, in light of their toxicity or other potentiality for harmful effect, or the method of their use, or the collateral measures necessary to use, they are not safe for use except under the supervision of a practitioner licensed by law to administer them. Indeed, all aromatase inhibitors that have been approved for marketing by the FDA are limited by an approved new drug application to use under the professional supervision of a practitioner licensed by law to administer such drug.


According to section 502(f)(1) of the Act, 21 U.S.C. § 352(f)(1), a drug is misbranded if, among other things, it fails to bear adequate directions for its intended use(s). Under 21 C.F.R. § 201.5, "adequate directions for use" means directions under which a layman can use a drug safely and for the purposes for which it is intended. Prescription drugs can only be used safely at the direction, and under the supervision, of a licensed practitioner. Therefore, it is impossible to write "adequate directions for use" for prescription drugs. FDA-approved drugs which bear their FDA-approved labeling are exempt from the requirement that they bear adequate directions for use by a layperson. See 21 C.F.R. §§ 201.100(c)(2) and 201.115. Because there are no FDA-approved applications for your firm's "Advanced Muscle Science Arom-X," "Advanced Muscle Science 4-AD UTT," "G.E.T ArimaDex," "iForce Nutrition Reversitol," and "Fizogen Off Cycle II Hardcore" products, their labeling fails to bear adequate directions for its intended uses, causing it to be misbranded under section 502(f)(1) of the Act, 21 U.S.C. § 352(t)(1). The introduction or delivery for introduction into interstate commerce of these misbranded products violates section 301(a) of the Act, 21 U.S.C. § 331(a).


The issues and violations cited in this letter are not intended to be an all-inclusive statement of the violations that exist in connection with your products. You are responsible for investigating and determining the causes of the violations identified above and for preventing their recurrence or the occurrence of other violations. It is your responsibility to ensure that all products marketed by your firm comply with the Act and its implementing regulations.


You should take prompt action to correct the violations cited in this letter. Failure to promptly correct these violations may result in legal action, without further notice, including, without limitation, seizure and injunction. Other federal agencies may take this Warning Letter into account when considering the award of contracts.


Within fifteen (15) working days of receipt of this letter, please notify this office in writing of the specific steps that you have taken to correct violations. Include an explanation of each step being taken to prevent the recurrence of violations, as well as copies of related documentation. If you cannot complete corrective action within fifteen working days, state the reason for the delay and the time within which you will complete the corrections. Furthermore, please advise this office what actions you will take to address product that you have already distributed.


Additionally, if another firm manufactures the products identified above, your reply should include the name and address of the manufacturer. If the firm from which you receive the product is not the manufacturer, please include the name of your supplier in addition to the manufacturer.
 

Please send your reply to Derek C. Price, Compliance Officer, U.S. Food and Drug Administration, 60 Eighth Street, N.E., Atlanta, GA 30309. If you have any questions about the content of this letter, please contact Mr. Price at 404-253-2277.


Sincerely
/S/
John Gridley
District Director
Atlanta District Office

1 We note that in 2003, (b)(4) submitted a New Dietary Ingredient Notification (NDIN) for piracetam, naming your firm as the distributor, pursuant to section 413(a)(2) of the Act, 21 U.S.C. §350b. Section 413(a)(2) of the Act requires premarket notification to CFSAN prior to the introduction of a new dietary ingredient into interstate commerce. The NDIN is required to contain infonnation which is the basis for the conclusion that a dietary supplement containing such new dietary ingredient will reasonably expected to be safe. CFSAN's response letter
to Mr. (b)(4) NDIN, which was issued on January 9, 2004, stated that piracetam is "not a dietary ingredient."
 

-

Reddick Dairy 8/30/10

  

Department of Health and Human Services logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 

PHILADELPHIA DISTRICT

900 U.S. Customhouse
2nd and Chestnut Streets
Philadelphia, PA 19106
Telephone: 215-597-4390

WARNING LETTER
10-PHI-11


CERTIFIED MAIL
RETURN RECEIPT REQUESTED


August 30, 2010


Keith A. Reddick, Owner
Reddick Dairy
2705 Sandy Lake Grove City Road
Stoneboro, Pennsylvania 16153


Dear Mr. Reddick:


On March 23 - 24, 2010, the U.S. Food and Drug Administration (FDA) conducted an investigation of your dairy operation located at 2705 Sandy Lake Grove City Road, Stoneboro, Pennsylvania. This letter notifies you of the violations of the Federal Food, Drug, and Cosmetic Act (the Act) that we found during our investigation of your operation. You can find the Act and its associated regulations on the Internet through links on FDA's web page at www.fda.gov.


We found that you offered for sale an animal for slaughter as food that was adulterated. Under section 402(a)(2)(C)(ii) of the Act, 21 U.S.C. § 342(a)(2)(C)(ii), a food is deemed to be adulterated if it bears or contains a new animal drug that is unsafe under section 512 of the Act, 21 U.S.C. § 360b. Further, under section 402(a)(4) of the Act, 21 U.S.C, § 342(a)(4), a food is deemed to be adulterated if it has been held under insanitary conditions whereby it may have been rendered injurious to health.


Specifically, our investigation revealed that on or about August 17, 2009, you consigned a bob veal calf, identified with sales tag (b)(4) for slaughter as food. On or about August 18, 2009, (b)(4) slaughtered this animal. United States Department of Agriculture, Food Safety and Inspection Service (USDA/FSIS) analysis of tissue samples collected from this animal identified the presence of flunixin residues of (b)(4) parts per million (ppm) in the liver, and ppm in the muscle. Flunixin is not approved for use in calves to be processed for veal. The presence of this drug in edible tissue from this animal causes the food to be adulterated within the meaning of section 402(a)(2)(C)(ii) of the Act, 21 U.S.C. § 342(a)(2)(C)(ii).


Our investigation also found that you hold animals under conditions that are so inadequate that medicated animals bearing potentially harmful drug residues are likely to enter the food supply. For example, you failed to maintain complete treatment records. Your records did not include the drug used, the dosage amount, the route of administration, and withdrawal times. Food from animals held under such conditions is adulterated within the meaning of section 402(a)(4) of the Act, 21 U.S.C. 342(a)(4).


The above is not intended to be an all inclusive list of violations. As a producer of animals offered for use as food, you are responsible for ensuring that your overall operation and the food you distribute is in compliance with the law.

You should take prompt action to correct the above violations described in this letter and to establish procedures to ensure that these violations do not recur. Failure to do so may result in regulatory action without further notice such as seizure and/or injunction.


You should notify this office in writing of the steps you have taken to bring your firm into compliance with the law within fifteen (15) working days of receiving this letter. Your response should include each step that has been taken or will be taken to correct the violations and prevent their recurrence. If corrective action cannot be completed within fifteen (15) working days of receiving this letter, state the reason for the delay and the time frame within which the corrections will be completed. Please include copies of any available documentation demonstrating that corrections have been made.
 

Your written response should be sent to Robin M. Rivers, Compliance Officer, U.S. Food and Drug Administration, U.S. Customhouse, Room 900, 200 Chestnut Street, Philadelphia, Pennsylvania 19106. If you have any questions about this letter, please contact Compliance Officer Robin M. Rivers at 215-717-3076 or via e-mail at Robin.Rivers@FDA.HHS.GOV.


Sincerely,

/S/
Kirk D. Sooter
District Director
Philadelphia District Office
 

-

Bristol Myers Squibb Holdings Pharma., Ltd.

  

Department of Health and Human Services logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 

466 Fernandez Juncos Avenue
Puerta De Tierra
San Juan, Puerto Rico 00901·3223


WARNING LETTER
10-SJN-WL-06

 

August 30, 2010

CERTIFIED MAIL RETURN
RECEIPT REQUESTED

 
Mr. Jim Cornelius
Chairman and CEO
Bristol-Myers Squibb Company
345 Park Avenue
New York, NY 10154
 

Dear Mr. Cornelius:
 

During our March 17-31, 2010 inspection of your manufacturing facility, Bristol-Myers Squibb Holding Pharma Ltd., located in Manati, Puerto Rico, investigators from the Food and Drug Administration (FDA) identified significant violations of Current Good Manufacturing Practice (CGMP) regulations for Finished Pharmaceuticals, Title 21, Code of Federal Regulations (C.F.R.), Parts 210 and 211. These violations cause your drug products to be adulterated within the meaning of section 501(a)(2)(B) of the Federal Food, Drug, and Cosmetic Act (the Act) [21 U.S.C. § 351(a)(2)(B)] in that the methods used in, or the facilities or controls used for, their manufacture, processing, packing, or holding do not conform to, or are not operated or administered in conformity with, CGMP.
 

We have reviewed your firm's response dated April 14, 2010, and note that it lacks sufficient corrective actions.
 

Specific violations observed during the inspection include, but are not limited, to the following:
 

1. Your firm has not established or followed appropriate written procedures designed to prevent microbiological contamination of drug products purporting to be sterile [21 C.F.R. § 211.113(b)]. For example,
 

a. The operators at your facility have repeatedly failed to comply with your procedures for aseptic operations. Specifically, your operators have been observed to not comply with Standard Operating Procedure (SOP) (b)(4) for ex. Operators did not follow SOP requirements pertaining to interventions into the Class 100 (ISO 5) zone. 

Our review indicates that there are on-going problems with your personnel failing to comply with procedures. In your response, you state that your SOPs are inadequate and do not reflect actual practices. While you commit to revising your SOPs, it is FDA's expectation that your firm promptly correct all deficient procedures to ensure employees do not continue improper practices.
 

To ensure quality of the drug product in aseptic operations, your personnel must employ strict discipline as they comply with adequate and appropriate procedures. We acknowledge your response stating that supervisors will monitor the performance of the aseptic operators on a daily basis. Please provide more information on the extent of this supervision. Also, your response is inadequate because the use of a log book to document supervisory observation of the aseptic filling process does not provide assurance of adequate supervision. Please provide a plan that evaluates your training program, specifically the program's effectiveness and your assurances of personnel compliance to aseptic processes prior to certification to work in an aseptic area.
 

b. Your firm failed to design and perform an adequate aseptic process simulation (i.e., media fill) based upon the same controls used for routine production. We note that the (b)(4) was observed (b)(4) during actual production. However, this intervention was performed only once during the media fill in 2009. 

In your response, you state that you have amended your SOP (b)(4) to ensure that the media fill simulation maintains alignment with routine aseptic filling processes.
 

During aseptic operations, it is critical that your Quality Control Unit (QCU) has reviewed routine aseptic filling processes prior to the preparation of the aseptic process simulation protocol and that the new media fills include all appropriate interventions.
 

This is a repeat observation from our 2005 and 2009 inspection.
 

2. Your firm has not thoroughly investigated the failure of a batch or any of its components to meet its specifications whether or not the batch has already been distributed [21 C.F.R. § 211.192]. 

For example, your firm has failed to thoroughly investigate the recurrence of environmental excursions in your facility and has failed to verify the effectiveness of prior corrective actions that addressed similar environmental excursions. Specifically, you reported (b)(4) environmental samples, described as, (b)(4) from January to February 2010 for the Class 100 (ISO Grade 5) areas. These (b)(4) environmental excursions were not investigated by your firm.
 

Previously, you documented approximately (b)(4) environmental excursions (September 9 to November 19, 2009) as (b)(4) associated with (b)(4). As part of your investigation (b)(4) (opened on October 26, 2009), the activities and interventions conducted in the aseptic core area from September to November 2009, were evaluated. Your firm implemented corrective actions as a result, including personnel training in the areas of gowning inspection, aseptic techniques, and cleaning procedures.
 

In your response, you state that the environmental excursions were from non-product contact areas and had no adverse product impact. We disagree with this premise because many of the (b)(4) sample locations (e.g., (b)(4)) were in the class 100 (ISO Grade 5) room used to load and unload lyophilizers. Your data demonstrates that this area is prone to contamination by the aseptic area personnel loading and unloading the lyophilizers, which is a manual operation. Manual operations represent a higher risk and additional measures may be required to ensure that an appropriately controlled environment is utilized to manufacture drug products purporting to be sterile.
 

In addition, your investigation of gown quality does not provide assurance that your clean room gown supplier has implemented appropriate actions to prevent a reoccurrence of defective gowns being used in your clean rooms. It is your responsibility to ensure that your clean room gown supplier has adequate inspection procedures to remove damaged gowns from your clean room gown supply. There is no mention in your investigation of a decision to shorten the life span of gannents, supplied to your firm, to prevent the excessive wear found during your investigation.
 

3. Your firm has not established scientifically sound and appropriate specifications, standards, sampling plans, and test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality, and purity [21 C.F.R. § 211.160(b)].
 

For example, your Firm established acceptance criteria for major defects in the control Form entitled, "Working Inspection Qualification, WIQ," using (b)(4). The acceptance criteria is inadequate because your QCD approved the specification without adequate justification or a scientifically sound statistical analysis.
 

Your response is inadequate because you fail to provide a sound scientific rationale for either establishing your acceptance criteria or classifying defects upon visual inspection of your lyophilized products. Typically, vials, with glass particles and cracks, are considered critical defects. However, you have classified these defects as major defects without justification.
 

The violations cited in this letter are not intended to be an all-inclusive statement of the violations that exist at your facility. You are responsible for investigating and determining the causes of the violations identified above and for preventing their recurrence, as well as occurrence of other violations. It is your responsibility to ensure that your firm complies with all requirements of federal law and FDA regulations.
 

You should take prompt action to correct the violations cited in this letter. Failure to promptly correct these violations may result in legal action without further notice including, without limitation, seizure and injunction. Other federal agencies may take this Warning Letter into account when considering the award of contracts. Additionally, FDA may withhold approval of requests for export certificates, or approval of pending drug applications listing your facility, until the above violations are corrected. FDA may re-inspect to verify corrective actions have been completed.
 

Repeat citations from prior inspections indicate that your quality control unit is either not appropriately exercising its responsibilities or does not have the authority to carry out its responsibilities. Due to continuing CGMP issues at your firm, we recommend you engage a third party consultant having appropriate CGMP expertise to assess your firm's facilities, procedures, processes, and systems to ensure that your drug products have their appropriate identity, strength, quality, and purity.
 

Within fifteen working days of receipt of this letter, please notify this office in writing of the specific steps that you have taken to correct violations. Include an explanation of each step being taken to prevent the recurrence of violations and copies of supporting documentation. If you cannot complete corrective action within fifteen working days, state the reason for the delay and the date by which you will have completed the correction. Additionally, your response should state if you no longer manufacture or distribute the drug product(s) manufactured at this facility, and provide the date(s) and reason(s) you ceased production.
 

Your reply should be sent to the following address: Food and Drug Administration, Attention: Margarita Santiago, Compliance Officer, 466 Fernandez Juncos Avenue, San Juan, Puerto Rico 00901-3223. If you have any questions concerning the violations noted please contact Ms. Santiago at (787) 474-4789 or by electronic mail at margarita.santiago@fda.hhs.gov.
 

Sincerely,
/S/
Maridalia Torres
District Director
San Juan District
 

cc: Mrs. Ivonne Lassalle
Vice-President & General Manager
Bristol-Myers Squibb Holding Pharma, Ltd
P.O. Box 30100
Manati, PR 00674-3000
 

Ms. Roberta L. McKee, Senior Vice President Americas, Asia Pacific Drug Products
Operations and Ms. Donna Gulbinski, Senior Vice President Worldwide Quality &
Compliance
P.O. Box 191
New Brunswick, NJ 08903-0191

-

Friday, August 27, 2010

Matthew N. Songer, MD

  

Department of Health and Human Services logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 10903 New Hampshire Ave
Silver Spring, MD 20993-0002
 

WARNING LETTER

AUG 27 2010

VIA UPS EXPRESS

Matthew N. Songer, M.D.
Orthopaedic Surgery Associates
1414 W. Fair Avenue, Suite 190
Marquette, Michigan 49855

Dear Dr. Songer:

This Warning Letter is to inform you of objectionable conditions observed during the Food and Drug Administration (FDA) inspection conducted at your clinical site from April 19 to April 22, 2010 by an investigator from the FDA Detroit District Office. The purpose of this inspection was to determine whether your activities and procedures related to your participation in the clinical study titled "A Prospective, Multi-Center, Randomized, Controlled Clinical Trial Evaluating the Safety and Effectiveness of NuBac® Disc Arthroplasty," Investigational Device Exemption (IDE) (b)(4),complied with applicable federal regulations. The NuBac® Disc Arthroplasty System is a device as that term is defined in section 201 (h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321 (h). This letter also requests prompt corrective action to address the violations cited.

The inspection was conducted under a program designed to ensure that data and information contained in requests for IDEs, Premarket Approval (PMA) applications, and Premarket Notification submissions (510(k)) are scientifically valid and accurate. Another objective of the program is to ensure that human subjects are protected from undue hazard or risk during the course of scientific investigations.

Our review of the inspection report prepared by the district office revealed several violations of Title 21, Code of Federal Regulations (21 CFR) Part 812 Investigational Device Exemptions and Part 50 - Protection of Human Subjects. At the close of the inspection, the FDA investigator presented an inspectional observations Form FDA 483 for your review and discussed the observations listed on the form with you, Dale A. Moilanen, Chief Operating Officer, Orthopaedic Surgery Associates of Marquette (OSAM), (b)(6), RN, Nurse Coordinator and Clinical Staff Supervisor, and (b)(6) NP, Research Coordinator. The deviations noted on the Form FDA 483 and our subsequent review of the inspection report are discussed below:

1. Failure to include all elements of informed consent. [21 CFR 50.25(a) and 50.25(b)]

An investigator is responsible for ensuring that all the basic elements of informed consent are provided to each subject when seeking informed consent. Further, the investigator is responsible for ensuring that the informed consent document (ICD) include certain additional elements when appropriate. You have failed to adhere to the above-stated regulations. Examples of your failure include, but are not limited to the following:

All (b)(4) subjects who enrolled in the NuBac® Disc Arthroplasty study at your clinical site signed the ICD "Pioneer NUBAC: MGH: November 11, 2008." This ICD does not include the following required information:

• a statement that the study involves research;

• identification of any procedures which are experimental;

• a statement that the particular treatment or procedure may involve risks to the subject (or to the embryo or fetus, if the subject is or may become pregnant) which are currently unforeseeable (the investigational plan requires a series of x-rays and Magnetic Resonance Imaging or Computed Tomography evaluations which could be harmful to an embryo or fetus);

• a statement that participation is voluntary, that refusal to participate will involve no penalty or loss of benefits to which the subject is otherwise entitled, and that the subject may discontinue participation at any time without penalty or loss of benefits to which the subject is otherwise entitled; and

• an explanation of whom to contact for answers to pertinent questions about the research and research subjects' rights.

In your response, please provide copies of the following: a revised IRB approved ICD that includes all the elements as required by 21 CFR 50.25, documentation of Institutional Review Board (IRB) review and approval of the ICD, and your plan to adequately inform all of the subjects who are receiving experimental treatment.

2. Failure to conduct the investigation according to the signed agreement with the sponsor, the investigational plan, 21 CFR part 812, other applicable FDA regulations, and any conditions of approval imposed by an Institutional Review Board (IRB) or FDA. [21 CFR 812.110(b)]

A clinical investigator is responsible for conducting an investigation according to the signed agreement with the sponsor, the investigational plan, 21 CFR part 812, other applicable FDA regulations and any conditions of approval imposed by an IRB or FDA. In addition to your failure to conduct the investigation in accordance with applicable FDA regulations in 21 CFR part 50 identified above, you failed to conduct the investigation in accordance with the investigational plan and the signed agreement. Examples of your failure include, but are not limited to the following:

The Randomized NuBac® investigational plan requires the patient to follow a clinical study schedule, which includes a series of radiological studies (i.e., Films and Scans) performed within the following time windows: Pre-op (i.e., baseline), discharge, 6 weeks, 3 months, 6 months, 12 months, 24 months and annually thereafter. These films include: Anterior-Posterior (AP), lateral, flexion and extension x-rays; and at least one of the following scans: Computed Tomography (CT) or Magnetic Resonance Imaging (MRI). The schedule of radiographic studies and exams in the protocol specifies that the pre-op films and scan(s) are to be performed within (b)(4) of surgery. Please note that radiographs taken more than (b)(4) prior to surgery may not represent the subject's true clinical condition prior to surgery which may result in subjects inappropriately being included in the study. You did not take pre-operative films or scan(s) as required by the protocol for seven (7) out of (b)(4) subjects enrolled in the Randomized NuBac® study at your clinical site.

• Subject (b)(6) received the pre-op MRI on August 25, 2008, followed by surgery on February 11, 2009. The MRI was taken over five (5) months prior to implantation of the investigational device.

• Subject (b)(6) received the pre-op MRI on April 4, 2008, followed by surgery on March 4, 2009. The MRI was taken 11 months prior to implantation of the investigational device.

• Subject (b)(6) received the pre-op MRI on October 22, 2008, followed by surgery on March 11, 2009. The MRI was taken over four (4) months prior to implantation of the investigational device. Additionally, there is no documentation for both the pre-op Flexion and Extension X-rays as required by the protocol.

• Subject (b)(6) received the pre-op MRI on December 4, 2006, followed by surgery on April 22, 2009. The MRI was taken approximately two (2) years and five (5) months prior to implantation of the investigational device.

• Subject (b)(6) received the pre-op MRI on February 25, 2009, followed by surgery on July 1, 2009. The MRI was taken over four (4) months prior to implantation of the investigational device. Additionally, there is no documentation for both the pre-op Flexion and Extension X-rays as required by the protocol.

• Subject (b)(6) received the pre-op MRI on June 25, 2009, followed by surgery on December 17, 2009. The MRI was taken over five (5) months prior to implantation of the investigational device.

• Subject (b)(6) received the pre-op MRI and pre-op AP and Lateral X-rays on July 27, 2009, followed by surgery on February 1, 2010. The MRI and X-rays were taken over six (6) months prior to implantation of the investigational device. Additionally, there is no documentation for both the pre-op Flexion and Extension X-rays as required by the protocol.

Please provide information on what actions you plan to take to prevent the above deviations from recurring in the future. If you develop standard operating procedures, then submit a copy of these procedures, along with documentation demonstrating staff training in these procedures, when you respond to this letter.

3. Failure to maintain accurate, complete, and current records related to your participation in the investigation. [21 CFR 812.140(a)(1), 812.140(a)(3)(ii) and 812.140(a)(4)]

A clinical investigator shall maintain accurate, complete and current records relating to the investigator's participation in an investigation. These records include all correspondence with the IRS, each Subject's case history and exposure to the device, all relevant observations, including data and information on the condition of each subject upon entering and during the course of the investigation, the current protocol for the study of the investigational device, and documentation showing the dates of and reasons for each deviation from the
protocol. Examples of your failure to adhere to the above stated regulations include, but are not limited to the following:

• There is no documentation for IRS review and approval for the following: the study guide dated February 2009, the investigational plan referenced in this study guide, and the revised investigational plan dated December 22, 2008.

• The June 2,2009 amended protocol was approved by the IRS on August 12, 2009; yet your site did not maintain it.

• Exclusion criteria # 25 of the investigational plan includes pregnancy or interested in becoming pregnant within the next two years. There is no documentation about how this exclusion criterion was met. For example, there is no documentation of negative pregnancy testing or birth control consultation for the following child-bearing age female subjects: Subject (b)(6)(33 year old female) and subject (b)(6) (37 year old female).

We acknowledge that the sponsor's representative provided a copy of the amended protocol to your site on April 21, 2010, during the FDA inspection. An investigator is responsible for ensuring that an investigation is conducted according to the signed agreement, the investigational plan, and applicable FDA regulations to protect the rights, safety, and welfare of subjects and for the control of devices under investigation [21 CFR 812.100]. Please provide information on what actions you plan to take to ensure that all study documentation, including protocols, subject case histories, test results, and correspondence between your site, the IRB, and sponsor are maintained in order to prevent the above deviations from recurring in the future. If you develop standard operating procedures, please submit with your response, a copy of these procedures, along with documentation demonstrating staff training in these procedures.

Please note that there was no documentation of financial disclosure for Dr. (b)(6) at your clinical site; however, Dr. (b)(6). implanted the investigational device in one subject and was listed as the doctor to contact on the ICD. We acknowledge that Dr. (b)(6). provided a signed financial disclosure statement and curriculum vitae prior to the FDA inspection close-out meeting.

In addition to the violations described above, we have concerns about a 2009 video titled UMotion Preservation: Understanding & Treating Spinal Pain" that was made available through your office at OSAM and on your website, www.upbonedoctors.com. You stated in the video that "Artificial Disc is a new technology that allows for movement. ...The next wave of advancement in the treatment of the spine is a minimally invasive artificial disc and that is where the NuBac comes in... with the NuBac, we're able to preserve the patient's normal movement. If they still need a total disc at some point, the more invasive procedure, we can still do that; if they still need a fusion in the future, we can still do that. So you don't burn any bridges whereas once you fuse somebody, that bridge is burned - you never can go back." The video also presents a testimonial from a 2005 NuBac recipient who claims that she has had no complications from the surgery or placement of the device. In addition, in the video you compare NuBac with other devices and procedures, claiming reduction in surgery time and quicker recovery. Although the video contains a brief written disclaimer at the bottom of the screen which states that the device is investigational, the claims made in the video suggest that NuBac is safe and effective and the video does not describe any of the potential risks associated with this investigational device.

The claims made in the video and the omission of risk information could unduly influence subjects to participate. Since the inspection, we have reviewed your website and acknowledge that you have removed the video from your website. FDA believes it is important that materials which could be viewed by prospective subjects are accurate and include appropriate risk information. In your response, please provide information on any actions that you plan to take to make sure that any such materials in the future are accurate and include appropriate risk information.

The violations described above are not intended to be an all inclusive list of problems that may exist with your clinical study. It is your responsibility as a clinical investigator to ensure compliance with the Act and applicable regulations.

Within fifteen (15) working days of receiving this letter, please provide written documentation of the actions you have taken or will take to correct these violations and prevent the recurrence of similar violations in current or future studies for which you are the clinical investigator. Any submitted corrective action plan must include projected completion dates for each action to be accomplished. Failure to respond to this letter and take appropriate corrective action could result in the FDA taking regulatory action without further notice to you. In addition, FDA could initiate disqualification proceedings against you in accordance with 21 CFR 812.119.

You will find information to assist you in understanding your responsibilities and planning your corrective actions in the FDA Information Sheets Guidance for Institutional Review Boards and Clinical Investigators. which can be found at http://www.fda.gov/oc/ohrtlirbs/.

Your response should reference "CTS # (b)(4)" and be sent to:

Attention: Linda D. Godfrey
Food and Drug Administration
Center for Devices and Radiological Health
Office of Compliance
Division of Bioresearch Monitoring
10903 New Hampshire Avenue
Building 66, Room 3462
Silver Spring, Maryland 20993-0002

A copy of this letter has been sent to the FDA Detroit District Office, 300 River Place, Suite 500, Detroit, MI 48207. Please send a copy of your response to that office.

If you have any questions, please contact Linda D. Godfrey at (301) 796-5654 or Linda.Godfrey@fda.hhs.gov.
 

Sincerely yours,
/S/
Larry D. Spears
Deputy Director of Regulatory Affairs
Office of Compliance
Center for Devices and
Radiological Health
 

cc:
Jeff Millen                    SPONSOR
President and CEO
Pioneer Surgical Technology, Inc.
375 River Park Circle
Marquette, Michigan 49855
 

James Surrell, M.D.                    IRB  
Marquette General Health System IRB
580 W College Avenue
Marquette, Michigan 49855
 

-

Wednesday, August 25, 2010

Flagship Niagara League Inc. 8/25/10

  

Department of Health and Human Services' logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 PHILADELPHIA DISTRICT
900 U.S. Customhouse
2nd and Chestnut Streets
Philadelphia, PA 19106
Telephone: 215-597-4390

WARNING LETTER
10-PHI-10

DELIVERED BY HAND

August 25, 2010

Wesley W. Heerssen, Jr., Master
Flagship Niagara League, Inc.
150 E. Front Street
Erie, Pennsylvania 16507

Dear Mr. Heerssen:

In July 2010, Erie County Department of Health officials reported an outbreak of Salmonella serotype Enteritidis (SE) on board your passenger vessel, Flagship Niagara, when the vessel was located near Cleveland, OH on July 10-11, 2010. Their investigation of the outbreak concluded that the ill individuals ate food items that were served on board your vessel, including a Mexican chicken diner and macaroni salad that contributed to the outbreak.

Subsequently, on July 17 and 19, 2010, the U.S. Food and Drug Administration (FDA) conducted an inspection of your passenger vessel, located at that time at 150 E. Front Street, Erie, Pennsylvania, which operates in interstate waters between its home port in Erie, Pennsylvania and various locations in the Great Lakes, including Cleveland, Ohio, Chicago, Illinois, and Duluth, Minnesota. This comprehensive inspection was conducted under the authority of the Public Health Service Act (PHS Act) to determine your compliance with applicable sections of the Interstate Conveyance Sanitation regulations (21 CFR Part 1250). These regulations were promulgated pursuant to Section 361 of the PHS Act (42 U.S.C. 264). Regulations promulgated under this section are necessary to prevent the introduction, transmission, or spread of communicable diseases. This inspection revealed significant deviations from 21 CFR Part 1250, under the PHS Act.

At the conclusion of the inspection, the FDA investigator issued a list of Inspectional Observations (Form FDA-483) to you. As documented on this form (copy enclosed), the following are the significant violations noted on your passenger vessel:

1. Your passenger vessel does not have water available at adequate temperatures, to all rooms in which food is prepared and utensils are cleaned, as required by 21 CFR 1250.30(c). Specifically, there is no hot water available on your ship for hand washing and cleaning purposes.

2. Your passenger vessel does not have a refrigerator that is equipped with a thermometer located in the warmest portion thereof, as required by 21 CFR 1250.34. Specifically, a thermometer was not installed in the (b)(4) brand reach-in refrigerator located in the ship's galley.

3. Your passenger vessel does not store all utensils used in connection with the preparation, storage, and serving of food or beverages so as to prevent contamination before reuse, as required by 21 CFR 1250.33(c). Specifically, clean pots and pans were stored on the floor of the ship's galley. Storage of these items directly on the floor of the galley does not adequately prevent contamination of these items before use.

The inspectional observations, identified above, are not intended to be an all-inclusive list of the conditions observed at your facility. It is your responsibility to assure adherence with all requirements of the regulations at this facility, and any other facilities involving interstate travel and sanitation under your control.

We may take further action if you do not promptly correct these violations. For instance, we may take further action to enjoin your passenger vessel from operating. It is your responsibility to ensure that all requirements of the PHS Act, the Federal Food, Drug, and Cosmetic Act, and their implementing regulations, are being met.

In addition, FDA recommends that the person in charge of your vessel, and your ship's cook, be trained in safe food-handling practices, such as properly cooking and cooling potentially hazardous food.

Please notify this office in writing within fifteen (15) working days from your receipt of this letter of the specific steps you have taken to correct the violations described above. Include in your response documentation to show correction has been achieved. If you cannot complete all corrections before you respond, state the reason for the delay and when you will complete the corrections.

Your response should be sent to Lynn S. Bonner, Compliance Officer, at the address noted above. If you have any questions with regard to this letter, Ms. Bonner can be reached at 215-717-3074 or Lynn.Bonner@fda.hhs.gov.

Sincerely,

/s/

Kirk D. Sooter
District Director
Philadelphia District Office


Enclosure: Form FDA 483
 

-

Activa Brand Products, Inc. 8/25/10

  

Department of Health and Human Services logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 10903 New Hampshire Avenue
Silver Spring, MD 20993 

AUG 25 2010

WARNING LETTER


VIA UNITED PARCEL SERVICES


Mr. Kirk Foley
President
Activa Brand Products, Inc.
112 W.B. MacPhail Dr.
Cornwall,
Canada C0A 1H0


Dear Mr. Foley:


During an inspection of your firm located in Cornwall, Canada on May 25 through May 27, 2010, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures AdvantaJet, AdvantaJetES, and GentleJet Needle-Free Injectors. Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. 321 (h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or are intended to affect the structure or function of the body.


This inspection revealed that these devices are adulterated within the meaning of section 501(h) of the Act (21 U.S.C. § 351(h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation found at Title 21, Code of Federal Regulations (C.F.R.), Part 820. We received a response from Mr. Gary Edwards, Director of Quality, dated June 14,20 10, concerning our investigator's observations noted on the Form FDA 483, List of Inspectional Observations that was issued to you. We address this response below, in relation to each of the noted violations. These violations include, but are not limited to, the following:


1. Failure to establish and maintain adequate procedures to ensure that where the results of a process cannot be fully verified by subsequent inspection and test, the process shall be validated with a high degree of assurance and approved, as required by 21 CFR 820.75(a).


For example, you did not provide any validation testing data to affirm that the packaging, (b)(4) + Tyvek packaging sealed with the (b)(4) was constructed to protect the vial adapter from damage during the conditions of processing, storage, handling, and distribution. Additionally, during conversation with the FDA investigator, you stated that you had not proceeded with the validation.


The adequacy of your response dated June 14, 2010, can not be determined at this time. You stated that your firm would initiate the actions (seal strength test, dye penetration test, and microbial barrier test) identified in a letter to CDRH dated December 3, 2008. You stated that you intend to have all these tests and validations completed by July 30, 2010. You should provide CDRH with the completed validation reports in your response to this letter.


2. Failure to establish and maintain adequate procedures for implementing corrective and preventive action, as required by 21 CFR 820.100(a).


For example, six(6) of the twenty(20) AdvantaJet injector service repair reports reviewed by the investigator required replacement of coiled springs from lot (b)(4) due to flaking of the outer coating. The flaking interfered with the bearings of the device and caused the device to "lock-up" and not deliver insulin. Your procedure, "Corrective and Preventive Action," Section QAM/P14, requires that a Corrective and Preventive Action Request form (ACT 007) be initiated when a need for corrective or preventive actions is identified in the repair reports. You stated that  (b)(4)injectors from lot (b)(4) had issues due to flaking of the coiled spring. After these failures were known, your firm did not analyze the repair reports adequately enough to identify potential causes of failure and a need for corrective action. Specifically, your firm's manager verbally identified the corrective action as rework of the coiled springs by sandblasting and buffing to remove the coating. However, no documentation of your firm's investigation into the failure, the identification of the corrective action, the validation of the effectiveness of the corrective action or the implementation of the corrective action was available. Additionally, all of the corrective or preventive actions and their effectiveness were to be recorded according to the procedure. However, for the defective coiled springs identified as Batch (b)(4) none of these activities were documented as required by the procedure.


The adequacy of your response dated June 14, 2010, cannot be determined at this time because the investigation into the coiled spring failure is ongoing. You stated that CAPA #(b)(4) was initiated to review the coil spring issue and determine why no corrective actions were initiated after this issue was discovered. You also stated that a full investigation and risk assessment were being conducted to determine if a recall of lot (b)(4) is necessary. You stated that the investigation and risk assessment should be completed by July 30, 2010. You should provide CDRH with the conclusion of your investigation into whether or not a recall of lot (b)(4) is necessary, per 21 CFR 806, as soon as possible.


3. Failure to ensure that each production run, lot, or batch of finished devices meets acceptance criteria prior to being released, as required by 21 CFR 820.80(d).


For example, your firm released an AdvantaJet (serial #(b)(4)) injector for distribution after the product's (b)(4) unit calibration was tested and failed to meet the acceptance criteria. According to the "Testing and Final Inspection of Injectors" procedure (Section ACT WI 002 Rev. A), if the calibration test does not pass the initial test, the injector is re-set, re-calibrated, re-tested, and the
results recorded on ACT 005 (Calibration and Penetration Test Sheet). Specifically, ACT 005 for serial # (b)(4) indicates that the first test failed at (b)(4) units with a measurement of (b)(4), but, there is no indication that a repeat test was performed because the "Test 2" box next to the"recording is empty. Because there is no repeat test score documented on the form, it can not be determined if the injector met the predetermined acceptance criteria prior to being released for distribution.


Your response dated June 14, 2010, is not adequate. Your response does not address or provide documentation to demonstrate why this product was released for distribution without meeting the predetermined acceptance criteria. Also, you stated that form ACT 005 (Calibration and Penetration Test Sheet) was updated to specifically require that all tests performed be documented on the form. But, after review of your revised ACT 005 form, it can not be determined how the new updates will require and instruct the user to document all tests performed. Because you did not provide a copy of the "Testing and Final Inspection of Injectors" procedure in your response, it is unclear if this procedure was updated to reflect the changes made to ACT 005.


Given the serious nature of the violations of the Act, AdvantaJet, AdvantaJetES, and GentleJet Needle-Free Injectors manufactured by your firm are still subject to refusal of admission under section 801(a) of the Act, 21 U.S.C. § 381(a), in that they appear to be adulterated. The FDA took steps on August 15, 2007, to refuse your products, known as "detention without physical examination," until violations are corrected. Your firm will remain on import alert. In order to remove the devices from detention, you should provide a written response to this Warning Letter as described below and correct the violations described in this letter. We will notify you if your response is adequate, and we may need to re-inspect your facility to verify that the appropriate corrections have been made.


Also, U.S. federal agencies are advised of the issuance of all Warning Letters about devices so that they may take this information into account when considering the award of contracts. Additionally, premarket approval applications for Class III devices to which the Quality System regulation deviations are reasonably related will not be approved until the violations have been corrected. Requests for Certificates to Foreign Governments will not be granted until the violations related to the subject devices have been corrected.


Please notify this office in writing within fifteen (15) working days from the date you receive this letter of the specific steps you have taken to correct the noted violations, including an explanation of how you plan to prevent these violations, or similar violations, from occurring again. Include documentation of the corrective action you have taken. If your planned corrections will occur over time, please include a timetable for implementation of those corrections. If corrective action cannot be completed within 15 working days, state the reason for the delay and the time within which the corrections will be completed. Please provide a translation of documentation not in English to facilitate our review.


Your response should be sent to:


U.S. Food and Drug Administration
Center for Devices and Radiological Health
ATTN: Valerie A. Flournoy WO66 Rm. 3526
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002


If you have any questions about the content of this letter please contact: LT John W. Diehl at (301) 796-0993 or (301) 847-8137 (fax).


Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your facility. It is your responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the Inspectional Observations, Form FDA 483 (FDA 483), issued at the closeout of the inspection may be symptomatic of serious problems in your firm's manufacturing and quality assurance systems. You should investigate and determine the causes of the violations, and take prompt actions to correct the violations and to bring your products into compliance.
Sincerely yours,

/S/

Timothy A. Ulatowski
Director
Office of Compliance
Center for Devices and
Radiological Health

-

Tuesday, August 24, 2010

Storz Medical, AG 8/24/10

 

Department of Health and Human Services logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 10903 New Hampshire Avenue
Silver Spring, MD 20993 

AUG 24 2010

WARNING LETTER


VIA UPS EXPRESS


Mr. Gerold Heine
Managing Director
Storz Medical, AG
Lohstampfestrasse 8
CH-8274 Tagerwilen
Switzerland


Dear Mr. Heine:


During an inspection of your firm located in Tagerwilen, Switzerland, on March 15, 2010, through March 19, 2010, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures the Modulith SLK and Modulith SLX-F2 devices. Under section 201(h) of the Federal Food, Drug and Cosmetic Act (the Act), 21 U.S.C. 321 (h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment or prevention of disease, or are intended to affect the structure or function of the body.


This inspection revealed that these devices are adulterated within the meaning of section 501 (h) of the Act (21 U.S.C. § 351 (h)), in that the methods used in, or the facilities or controls used for, their manufacture, packing. storage, or installation are not in conformity with the Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation found at Title 21, Code of federal Regulations (C.F.R.), Part 820. We received a response from Klaus Hugen, Director of Quality Management, dated March 29, 2010, concerning our investigator's observations noted on the Form FDA 483, List of Inspectional Observations that was issued to you. We acknowledge receipt of the additional response dated April 26, 2010, however, this response was received after fifteen days of the close of the inspection and as a result, was not reviewed. We address the March 29, 2010, response below, in relation to each of the noted violations. These violations include, but are not limited to, the following:


1. Failure to establish and maintain adequate procedures for verifying or validating the corrective and preventive action to ensure that such action is effective and does not adversely affect the finished device, as required by 21 CFR 820.100(a)(4). For example, no protocol, including acceptance criteria, was established for the validation of Change Request (b)(4). Additionally, there was no documentation showing that this change was validated. The change was implemented to fix cracked cooling pumps in the Modulith SLX-F2.


We reviewed your response and conclude that it is inadequate because you have not provided evidence of a correction, corrective action, or systemic corrective action. Your response indicates your intention to revise the procedure, Design and Technical Changes, and provide training on the procedure. You did not provide evidence of this revision or evidence of a systemic corrective action.


2. failure to establish and maintain adequate procedures to control product that does not meet specified requirements, as required by 21 CFR 820.90(a). For example, your firm failed to follow procedure QMV_SMAG_002_01_02, Control of Nonconforming Products, in that there was no documentation of the evaluation and disposition of the following pumps found defective ("intake cracked") during production: Serial numbers (b)(4); (b)(4) and (b)(4)

We reviewed your response and conclude that that it is inadequate because you did not provide evidence of a correction, corrective action, or systemic corrective action. You indicate in your response that you will retrain employees on the onconforming Products procedure, but you have not provided evidence of this retraining or evidence of any additional corrective action or systemic corrective action.


3. Failure to establish and maintain adequate procedures to ensure that any complaint involving the possible failure of a device, labeling, or packaging to meet any of its specifications shall be reviewed, evaluated, and investigated unless such investigation has already been performed for a similar complaint and another investigation is not necessary, as required by 21 CFR 820.198(c). For example, your firm failed to extend the investigation into the causes of Complaint# (b)(4) to additional units exhibiting the same defect reported in the complaint. This complaint was opened on 11/14/09, following the receipt of Product Complaint Report# (b)(4) describing a crack at the inlet hose attachment of the cooling pump in Modulith SLX-F2 system. According to the complaint the system "was noticed to be leaking water." The complaint was investigated and closed on 3/15/10. During the complaint investigation three pumps with the same defect were detected in production. However, there is no documentation showing that these three units were evaluated and results or conclusions included or referred to in Complaint# (b)(4)


We reviewed your response and conclude that it is inadequate because you have not provided evidence of a correction, corrective action, or systemic corrective action. You indicate in your response the intention to revise your complaint procedure and provide training. Your firm indicated that a timeline regarding the nonconforming product in the observation was attached, but this could not be located within the response. Please provide evidence of the revised procedure, training, and any systemic corrective action that has been completed.

4. Failure to establish and maintain adequate procedures to ensure that the manufacturer shall review and evaluate all complaints to determine whether an investigation is necessary and maintain a record that justifies when no investigation is performed, as required by 21 CFR 820.198(b). For example, the firm failed to conduct an investigation into the causes of Complaint (b)(4). This complaint was opened on 5/8/09, after receiving Product Complaint Report# (b)(4) from a US distributor, Product Complaint Report# (b)(4) reported "water leaking from the cooling pump" and a crack in the pump inlet port in the Modulith SLX-F2 system. The defective pump was returned on 8/12/09 and the firm decided not to evaluate the complaint for investigation. Even without a documented investigation, including root cause, change request (b)(4) was initiated on 11/19/09 and approved on 1/18/10 to remediate the water leaking defect. Complaint (b)(4) was closed on 3/15/10. There is no documentation showing the reason not to investigate and the individual responsible for the decision.


We reviewed your response and conclude that it is inadequate because you have not provided evidence of a correction, corrective action, or systemic corrective action. In your response you acknowledge the need to revise your complaint procedure, but you did not provide this for review. You also did not provide any corrective action or systemic corrective action.


A follow up inspection will be required to assure that corrections arc adequate. An FDA trip planner will be in touch with you to arrange a mutually convenient date for this inspection.


You should take prompt action to correct the violations addressed in this letter. Failure to promptly correct these violations may result in regulatory action, which may include detaining your devices without physical examination upon entry into the United States until the corrections are completed. Section 801(a) of the Act (21 U.S.C. § 381(a)) Also, U.S. federal agencies are advised of the issuance of all Warning Letters about devices so that they may take this information into account when considering the award of contracts. Additionally, premarket approval applications for Class III devices to which the Quality System regulation deviations are reasonably related will not be approved until the violations have been corrected. Requests for Certificates to Foreign Governments will not be granted until the violations related to the subject devices have been corrected.


Please notify this office in writing within fifteen (15) working days from the date you receive this letter of the specific steps you have taken to correct the noted violations, including an explanation of how you plan to prevent these violations, or similar violations, from occurring again. Include documentation of the corrective action you have taken. If your planned corrections will occur over time, please include a timetable for implementation of those corrections. If corrective action cannot be completed within 15 working days, state the reason for the delay and the time within which the corrections will be completed. If the documentation is not in English, please provide a translation to facilitate our review.


Your response should be sent to: Matthew Krueger, WO-66 RM 3676, 10903 New Hampshire Avenue. Silver Spring, Maryland 20993. If you have any questions about the content of this letter please contact: Amy Skrzypchak at (301) 796-5613 or amy.skrzypchak@fda.hhs.gov.


Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your facility. It is your responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the Inspectional Observations, Form FDA 483 (FDA 483), issued at the closeout of the inspection may be symptomatic of serious problems in your firm's manufacturing and quality assurance systems. You should investigate and determine the causes of the violations, and take prompt actions to correct the violations and to bring your products into compliance.


Sincerely your,

/S/
Timothy A. Ulatowski
Director
Office of Compliance
Center for Devices and
Radiological Health

-

18th Street Deli, Inc

  

Department of Health and Human Services logoDepartment of Health and Human Services

Public Health Service
Food and Drug Administration
 Detroit District
300 River Place
Suite 5900
Detroit, MI 48207
Telephone: 313-393-8100
FAX 313-393-8139

 

WARNING LETTER
2010-DT-17
 
August 24, 2010
 
VIA UPS
 
David L. Salerno, President
18th Street Deli, Inc.
8800 Conant Street
Hamtramck, MI 48211-1401
                                                                                               
Dear Mr. Salerno:
 
We inspected your seafood processing facility, located at 8800 Conant Street, Hamtramck, MI 48211-1401 on April 1-2, 5 and 8, 2010. We found that you have serious violations of the seafood Hazard Analysis and Critical Control Point (HACCP) regulation, Title 21, Code of Federal Regulations, Part 123 (21 CFR 123). In accordance with 21 CFR 123.6(g), failure of a processor of fish or fishery products to have and implement a HACCP plan that complies with this section or otherwise operate in accordance with the requirements of Part 123, renders the fish or fishery products adulterated within the meaning of Section 402(a)(4) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 342(a)(4). 
 
Accordingly, your ready-to-eat tuna salad, tuna pasta salad, and fish and cheese sandwich products are adulterated, in that they have been prepared, packed, or held under insanitary conditions whereby they may have been rendered injurious to health. You may find the Act, the seafood HACCP regulation and the Fish and Fisheries Products Hazards & Controls Guidance through links in FDA's home page at www.fda.gov.
 
We acknowledge your response received on April 26, 2010 that included a revised HACCP plan and summaries of your corrections. We have provided comments concerning the information below.
 
Your significant violations were as follows:
 
1) You must conduct or have conducted a hazard analysis to determine whether there are food safety hazards that are reasonably likely to occur and have a HACCP plan that, at a minimum, lists the critical control points, to comply with 21 CFR 123.6(c)(2). A critical control point is defined in 21 CFR 123.3(b) as a "point, step, or procedure in a food process at which control can be applied and a food safety hazard can as a result be prevented, eliminated, or reduced to acceptable levels." However, your firm’s HACCP plan for “Ready to Eat Products” for fish and cheese sandwiches does not list a  critical control point at the “Refrigerated Staging for Shipping” step that is necessary for to control the food safety hazards of pathogens due to time/temperature abuse. Specifically, ready to eat cod and cheese sandwiches are routinely manufactured by your firm and are stored in the Staging/Pack Out Cooler for up to 48 hours prior to shipment to your customers.
 
Your revised HACCP plan provided with your response dated April 26, 2010 references “Refrigerated Staging” at the same critical control point for “refrigerated storage”. However, there is no information to distinguish between the two critical control points and no evidence that these will be monitored as individual critical control points
 
2) You must have a HACCP plan that, at a minimum, lists the food safety hazards that are reasonably likely to occur, to comply with 21 CFR 123.6(c)(1). A food safety hazard is defined in 21 CFR 123.3(f) as "any biological, chemical, or physical property that may cause a food to be unsafe for human consumption." However, your firm’s HACCP plans do not list the following hazards:
 
a) “Ready to Eat Products” does not list the food safety hazards of Allergens
 
b) “Tuna/Tuna Pasta Salad” does not list the food safety hazards of Allergens or Metal fragments.
 
Your response received on April 26, 2010 does not list the above hazards and does not provide any information to demonstrate control of these hazards identified in the plans above. 
 
3) You must implement the record keeping system that you listed in your HACCP plan, to comply with 21 CFR 123.6(b). However, your firm does not consistently maintain temperature monitoring records at the Refrigerated Storage critical control point to control pathogen growth and histamine formation as listed in your HACCP plan for Ready to Eat Tuna Salad and Tuna Pasta Salad. Specifically your firm had no temperature monitoring records for your Main Cooler for January 11 and 13, 2010, February 10, 15 and 17, 2010 while ready to eat tuna salad was being stored there.
 
4) You must have a HACCP plan that, at a minimum, lists monitoring procedures and their frequency for each critical control point, to comply with 21 CFR 123.6 (c)(4). However, your firm’s HACCP plan for “Tuna/Tuna Pasta Salad” lists a monitoring procedure/ frequency at the “Refrigerated Storage and Refrigerated Staging for Shipping” critical control point that is not adequate to control pathogen growth and histamine formation. FDA recommends that reports be reviewed on a daily basis to ensure that the Critical Limit is not exceeded and that the equipment is functioning properly.
 
We acknowledge that your response received on April 26, 2010, states that the Data Logger will be trended twice a month. Your firm further responded that you will review critical control point monitoring records weekly. We will verify this during the next inspection.
 
5) You must monitor sanitation conditions and practices during processing with sufficient frequency to ensure compliance with current good manufacturing practice requirements in 21 CFR Part 110, to comply with 21 CFR 123.11(b). However, during the manufacturing of tuna salad on April 5, 2010, your firm was using a white spatula that that had unsmooth surfaces and edges that were frayed, creviced and not easily cleanable and was in direct contact with mayonnaise and salad dressing that was used in the tuna salad. However, your sanitation monitoring records for that same day indicate that these food contact surfaces were acceptable.
 
6) You must maintain adequate sanitation control records that, at a minimum, document monitoring and corrections set out in 21 CFR 123.11(b), to comply with 21 CFR 123.11 (c). However your firm did not maintain adequate sanitation monitoring records for the condition of food contact surfaces required for the processing of tuna salad and/or tuna pasta salad. Specifically, seven of seven sanitation monitoring records reviewed lacked monitoring entries for the condition of food contact surfaces while tuna salad and/or tuna pasta salad was being manufactured on January 4 and 15, 2010, February 3 and 22, 2010,  March 5 and 17, 2010 and April 5, 2010. Also, six of the same seven sanitation monitoring records reviewed lacked entries for employee health monitoring. The records are inadequate because the requirements of 21 CFR 123.9(a)(2) and (a)(4) were not present.
 
We acknowledge that your response received April 26, 2010 states  that a new can opener and conveyor belt have been purchased and that the can opener has been added to the pre-operational sanitation inspection log. However; we do not have enough information to ascertain whether you are 1) monitoring sanitation conditions and practices during processing with sufficient frequency to ensure compliance with current good manufacturing practice requirements, 2) whether you are documenting monitoring and corrections as set out in 21 CFR 123.11(b), or 3) what corrective action was taken to address the spatula and other utensils that the investigator found “not easily cleanable”. 
 
We acknowledge you have included employee health conditions on your pre-operational sanitation inspection log in your response to the observation that you are not maintaining sanitation control records that document monitoring and corrections of sanitation deficiencies for condition and cleanliness of food contact surfaces and control of employee health conditions. However, your response is still inadequate in that your response did not include the condition of food contact surfaces to this inspection log. We request that you provide documentation that this inspection log has been implemented.
 
We may take further action if you do not promptly correct these violations. For instance, we may take further action to seize your product(s) and/or enjoin your firm from operating.
 
This letter may not list all the violations at your facility. You are responsible for ensuring that your processing plant operates in compliance with the Act, the seafood HACCP regulation (21 CFR Part 123) and the Current Good Manufacturing Practice regulation (21 CFR Part 110). You also have a responsibility to use procedures to prevent further violations of the Act and all applicable regulations.
 
You should respond in writing within fifteen (15) working days from your receipt of this letter. Your response should outline the specific things you are doing to correct these violations. You should include in your response documentation such as HACCP and verification records, or other useful information that would assist us in evaluating your corrections. If you cannot complete all corrections before you respond, you should explain the reason for your delay and state when you will correct any remaining violations.
 
Please send your reply to the Food and Drug Administration, Attention: Vanessa Colburn, Compliance Officer, 300 River Place, Suite 5900, Detroit, Michigan 48207.   If you have questions regarding any issues in this letter, please contact Ms. Colburn at 313-393-8121.
                                                                                
Sincerely,
/S/
Joann M. Givens
District Director
Detroit District Office
-